Features of construction :
- Sterilization chamber in SS 304 / SS 316 stainless steel with high quality
argon welding.
- Structure frame reinforcement for chamber to prevent thermal shocks and
vibrations.
- All joints fully welded and edges curved i.e. Crevice free working
chamber for hygine and easy cleaning
- Compacted glass wool to maintain uniform temp internally and reduced heat
losses.
Application
These equipments are suitable for sterilization of,
- Glassware like Ampoules, vials, etc.
- Metal trays, Metal material and containers.
The design philosophy of JF Dry Heat Sterilizer assures the improved
technique of reduction of bacterial level, sterilization and de-pyrogenation by
Dry Heat process for Pharma industries. These equipment have been designed to
work at a maximum working temperature upto 2500 C.
|
Doors
- Duly insulated doors with silicon gasket for positive sealing.
- Proper door locking with cam operate
mechanism
- Positive door interlocking of doors with pneumatically operated cylinders
to prevent simultaneous opening of both doors.
- Process door lock to prevent opening of doors when process is
'ON'.
Heat
Special tubular SS 304 / SS 316 air heaters for particles free operation.
Air circulation
- The efficiency of DRY HEAT process depends on velocity and the suitably
designed circulation pattern.
- The air flowing through the chamber is distributed uniformly with the help
of perforated wall having fixed opening.
- Specially designed motor mounting arrangement to lower down vibrations and
sounds level.
- Dynamically balanced SS 304 / SS 316 impellers.
- The pressure module with pre filter and HEPA filter to ensure fresh
filtered air entry - inside the chamber.
- HEPA filter located on exhaust to prevent any back flow of air that may
cause contamination of load. (Optional)
- Special air flow technique ensure over pressure throughout the cycle for
100 % sterile integrity.
|
 |
 |
Accessories
- Trolleys : SS 304 / SS 316 for transfer of carriages between working areas.
- Carriages : Fabricated from SS 304 / SS 316 and are designed to ensure
maximum penetration of the sterilizing medium with optimum utilization of
space.
- Cooling Coil : Fabricated from SS 304 / SS 316 finned tube to ensure rapid
cooling of product to desire temperature.
- Process Controls : JF Dry Heat Sterilizer are PLC / microprocessor
controlled. The microprocessor is easy to operate and offer automatic cycles,
digital readouts of all parameters. Circular / strip chart recorders, data
logger with printers. Semiautomatic control panel are also available.
- Also available in class 100 models.
|